Intraoperative navigation and preoperative templating software are associated with increased glenoid baseplate screw length and use of augmented baseplates in reverse total shoulder arthroplasty

Open AccessPublished:October 29, 2020DOI:https://doi.org/10.1016/j.jseint.2020.09.003

      Background

      Preoperative templating software and intraoperative navigation have the potential to impact baseplate augmentation utilization and increase screw length for baseplate fixation in reverse total shoulder arthroplasty (rTSA). We aimed to assess their impact on the (1) baseplate screw length, (2) number of screws used, and (3) frequency of augmented baseplate use in navigated rTSA.

      Methods

      We compared 51 patients who underwent navigated rTSA with 63 controls who underwent conventional rTSA at a single institution. Primary outcomes included the screw length, composite screw length, number of screws used, percentage of patients in whom 2 screws in total were used, and use of augmented baseplates.

      Results

      Navigation resulted in the use of significantly longer individual screws (36.7 mm vs. 30 mm, P < .0001), greater composite screw length (84 mm vs. 76 mm, P = .048), and fewer screws (2.5 ± 0.7 vs. 2.8 ± 1, P = .047), as well as an increased frequency of using 2 screws in total (35 of 51 patients [68.6%] vs. 32 of 63 controls [50.8%], P = .047). Preoperative templating resulted in more frequent augmented baseplate utilization (76.5% vs. 19.1%, P < .0001).

      Conclusion

      The difference in the screw length, number of screws used, and augmented baseplate use demonstrates the evolving role that computer navigation and preoperative templating play in surgical planning and the intraoperative technique for rTSA.

      Level of evidence

      Keywords

      Reverse total shoulder arthroplasty (rTSA) has become a mainstay of treatment for patients with rotator cuff arthropathy and is becoming more common for some patients with glenohumeral arthritis. The indications for rTSA are still evolving and are continuing to expand. In general, most available rTSA designs include a glenoid baseplate that is secured by a central peg, keel, or screw in the glenoid vault, with a variety of screw arrangement designs for supplemental fixation into scapular bony corridors. These supplemental screws can be locking or nonlocking and variable or fixed angle. Secure fixation of the baseplate is paramount to prevent glenoid component loosening.
      • Zumstein M.A.
      • Pinedo M.
      • Old J.
      • Boileau P.
      Problems, complications, reoperations, and revisions in reverse total shoulder arthroplasty: a systematic review.
      Inadequate initial glenoid baseplate fixation has been shown to lead to early glenoid loosening from micromotion and a lack of osseous integration.
      • Frankle M.
      • Siegal S.
      • Pupello D.
      • Saleem A.
      • Mighell M.
      • Vasey M.
      The reverse shoulder prosthesis for glenohumeral arthritis associated with severe rotator cuff deficiency. A minimum two-year follow-up study of sixty patients.
      Rates of baseplate loosening from 1.2 to 5% have been reported.
      • Bitzer A.
      • Rojas J.
      • Patten I.S.
      • Joseph J.
      • McFarland E.G.
      Incidence and risk factors for aseptic baseplate loosening of reverse total shoulder arthroplasty.
      ,
      • Zumstein M.A.
      • Pinedo M.
      • Old J.
      • Boileau P.
      Problems, complications, reoperations, and revisions in reverse total shoulder arthroplasty: a systematic review.
      Secure fixation of the baseplate to the glenoid has been attributed to a number of factors including bone quality, number of screws, screw length, arrangement and angle of divergence of screws, and central peg length. Numerous studies have attempted to examine the impact of these factors on fixation, with mixed results.
      • Chebli C.
      • Huber P.
      • Watling J.
      • Bertelsen A.
      • Bicknell R.T.
      • Matsen III, F.
      Factors affecting fixation of the glenoid component of a reverse total shoulder prothesis.
      ,
      • Hoenig M.P.
      • Loeffler B.
      • Brown S.
      • Peindl R.
      • Fleischli J.
      • Connor P.
      • et al.
      Reverse glenoid component fixation: is a posterior screw necessary?.
      ,
      • Humphrey C.S.
      • Kelly II, J.D.
      • Norris T.R.
      Optimizing glenosphere position and fixation in reverse shoulder arthroplasty, part two: the three-column concept.
      ,
      • Irlenbusch U.
      • Kohut G.
      Evaluation of a new baseplate in reverse total shoulder arthroplasty—comparison of biomechanical testing of stability with roentgenological follow up criteria.
      • James J.
      • Huffman K.R.
      • Werner F.W.
      • Sutton L.G.
      • Nanavati V.N.
      Does glenoid baseplate geometry affect its fixation in reverse shoulder arthroplasty?.
      • Kennon J.C.
      • Lu C.
      • McGee-Lawrence M.E.
      • Crosby L.A.
      Scapula fracture incidence in reverse total shoulder arthroplasty using screws above or below metaglene central cage: clinical and biomechanical outcomes.
      • Königshausen M.
      • Jettkant B.
      • Sverdlova N.
      • Ehlert C.
      • Gessmann J.
      • Schildhauer T.A.
      • et al.
      Influence of different peg length in glenoid bone loss: a biomechanical analysis regarding primary stability of the glenoid baseplate in reverse shoulder arthroplasty.
      • Lung T.S.
      • Cruickshank D.
      • Grant H.J.
      • Rainbow M.J.
      • Bryant T.J.
      • Bicknell R.T.
      Factors contributing to glenoid baseplate micromotion in reverse shoulder arthroplasty: a biomechanical study.
      However, it appears that, in general, longer screws that achieve bicortical purchase in dense bone provide the best fixation.
      • Humphrey C.S.
      • Kelly II, J.D.
      • Norris T.R.
      Optimizing glenosphere position and fixation in reverse shoulder arthroplasty, part two: the three-column concept.
      ,
      • James J.
      • Huffman K.R.
      • Werner F.W.
      • Sutton L.G.
      • Nanavati V.N.
      Does glenoid baseplate geometry affect its fixation in reverse shoulder arthroplasty?.
      ,
      • Lung T.S.
      • Cruickshank D.
      • Grant H.J.
      • Rainbow M.J.
      • Bryant T.J.
      • Bicknell R.T.
      Factors contributing to glenoid baseplate micromotion in reverse shoulder arthroplasty: a biomechanical study.
      Computer-assisted intraoperative navigation and preoperative planning software applications in rTSA have only recently been described in the orthopedic literature because of the relative infancy of the technology.
      • Parsons M.
      • Greene A.
      • Polakovic S.
      • Byram I.
      • Cheung E.
      • Jones R.
      • et al.
      Assessment of surgeon variability in preoperative planning of reverse total shoulder arthroplasty: a quantitative comparison of 49 cases planned by 9 surgeons.
      ,
      • Roche C.
      • DiGeorgio C.
      • Yegres J.
      • VanDeven J.
      • Stroud N.
      • Flurin P.H.
      • et al.
      Impact of screw length and screw quantity on reverse total shoulder arthroplasty glenoid fixation for 2 different sizes of glenoid baseplates.
      ,
      • Walch G.
      • Badet R.
      • Boulahia A.
      • Khoury A.
      Morphologic study of the glenoid in primary glenohumeral osteoarthritis.
      Patient-specific instrumentation and 3-dimensional (3D) imaging and templating technologies improve the accuracy of glenoid component placement in accordance with a preoperative plan.
      • Barrett I.
      • Ramakrishnan A.
      • Cheung E.
      Safety and efficacy of intraoperative computer-navigated versus non-navigated shoulder arthroplasty at a tertiary referral.
      ,
      • Berhouet J.
      • Gulotta L.V.
      • Dines D.M.
      • Craig E.
      • Warren R.F.
      • Choi D.
      • et al.
      Preoperative planning for accurate glenoid component positioning in reverse shoulder arthroplasty.
      ,
      • Iannotti J.P.
      • Weiner S.
      • Rodriguez E.
      • Subhas N.
      • Patterson T.E.
      • Jun B.J.
      • et al.
      Three-dimensional imaging and templating improve glenoid implant positioning.
      ,
      • Nguyen D.
      • Ferreira L.M.
      • Brownhill J.R.
      • King G.J.
      • Drosdowech D.S.
      • Faber K.J.
      • et al.
      Improved accuracy of computer assisted glenoid implantation in total shoulder arthroplasty: an in-vitro randomized controlled trial.
      ,
      • Verborgt O.
      • De Smedt T.
      • Vanhees M.
      • Clockaerts S.
      • Parizel P.M.
      • Van Glabbeek F.
      Accuracy of placement of the glenoid component in reversed shoulder arthroplasty with and without navigation.
      ,
      • Walch G.
      • Badet R.
      • Boulahia A.
      • Khoury A.
      Morphologic study of the glenoid in primary glenohumeral osteoarthritis.
      In addition, 3D templating may increase the accuracy of screw placement relative to the preoperative plan.
      • Venne G.
      • Rasquinha B.J.
      • Pichora D.
      • Ellis R.E.
      • Bicknell R.
      Comparing conventional and computer-assisted surgery baseplate and screw placement in reverse shoulder arthroplasty.
      To our knowledge, only 1 previous study has evaluated the difference in achievable baseplate screw length between navigated and conventional rTSA.
      • Nashikkar P.S.
      • Scholes C.J.
      • Haber M.D.
      Role of intraoperative navigation in the fixation of the glenoid component in reverse total shoulder arthroplasty: a clinical case-control study.
      We sought to evaluate the impact of preoperative templating software and intraoperative navigation on baseplate component specifications by comparing the (1) average screw length, (2) composite screw length (ie, combined length of all screws), (3) average number of screws used, (4) percentage of patients in whom 2 screws in total were used, and (5) frequency of augmented baseplate use. Operative time was examined as a secondary variable. We hypothesized that intraoperative navigation assistance would allow the primary surgeons to use longer and fewer screws to achieve what they deemed secure baseplate fixation, as compared with secure baseplate fixation when intraoperative navigation assistance was not used.

      Materials and methods

      Surgical technique

      All patients underwent rTSA with the Exactech Equinoxe system (Exactech, Gainesville, FL, USA), which has a design that uses a medialized glenoid and lateralized humerus.
      • Routman H.D.
      • Flurin P.H.
      • Wright T.W.
      • Zuckerman J.D.
      • Hamilton M.A.
      • Roche C.P.
      Reverse shoulder arthroplasty prosthesis design classification system.
      Augmented baseplates were available for use throughout the study period; these included baseplates with a superior, posterior, or superior-posterior augmentation, a small baseplate size, or an extended central peg. All standard-sized baseplates (including augmented baseplates) have 6 screw hole options (superior, inferior, superior-anterior, superior-posterior, inferior-anterior, and inferior-posterior), whereas the small baseplate has 4 screw hole options (superior, inferior, inferior-anterior, and inferior-posterior).
      In the large majority of non-navigated cases, the decision to use a standard or augmented glenoid baseplate component was made using a preoperative computed tomography (CT) scan (if available) combined with an intraoperative assessment of glenoid morphology. The Equinoxe Planning App preoperative planning software and ExactechGPS Total Shoulder Application intraoperative navigation system (Exactech) were used for all navigated cases (Fig. 1). This technology offers the surgeon a 3D rendered model of the patient’s scapula, created from the preoperative CT scan. The software then allows the surgeon to place sized-to-scale renderings of all baseplate options onto the scapula model and make adjustments in baseplate positioning in all planes in 1-mm and 1° increments. The presurgical plan is saved and uploaded to the intraoperative navigation system prior to surgery. Regardless of the use of navigation, the baseplate-glenoid interface was inspected intraoperatively to ensure a flush fit.
      Figure thumbnail gr1
      Figure 1The Equinoxe Planning App preoperative planning software is shown, demonstrating the ability to virtually construct the ideal glenoid baseplate type and position for each patient’s scapular anatomy. Post Aug, posterior augment; Post-Sup, posterosuperior augment; 10° Sup Aug, 10° superior augment; Anat., anatomic baseplate.
      A standard deltopectoral approach was performed with a 1- to 2-cm cranial extension of the incision for coracoid exposure in patients who underwent navigated rTSA. Navigation was not available for placement of the humeral component. Therefore, humeral head osteotomy, reaming, broaching, and placement of the humeral component were performed in similar fashions in both groups.
      In the non-navigated group, the baseplate central peg positioning and reaming of the glenoid were performed using a standard technique.
      Exactech
      Extremities operative technique: Equinoxe platform shoulder system.
      When navigation was used, the superior and inferior aspects of the coracoid were dissected with cautery to allow for placement of a tracker that corresponded with the intraoperative navigation screen. The tracker was then placed on the superior coracoid with 2 self-tapping screws. It was linked to the intraoperative monitor, and several points on the glenoid and coracoid were registered using a stylus. An additional tracker that corresponded with the intraoperative navigation screen was attached to the base of the ExactechGPS reamer handle, and the corresponding drills and reamers for each step in the procedure were attached to this reamer handle. Locating the center point of the glenoid, drilling of the center peg hole, and glenoid reaming were carried out under navigation, according to the preoperative plan. The selected baseplate component was then impacted onto the glenoid.
      The variable-angle compression screws were drilled using the adjustable-angle drill guide with a standard 3.2-mm drill bit for the non-navigated cases and with an ExactechGPS-specific drill bit for the navigated cases. The navigation screen displays the trajectory of the drill bit in relation to coronal and sagittal CT images of the bony scapula in real time, which allows the surgeon to aim in a manner that will access the longest bony corridor provided by each patient’s unique glenoid and scapular anatomy (Fig. 2). Pre-drill aiming also allows the surgeon to avoid areas of the scapula that could endanger soft-tissue structures, such as the suprascapular notch.
      • Hart N.D.
      • Clark J.C.
      • Wade Krause F.R.
      • Kissenberth M.J.
      • Bragg W.E.
      • Hawkins R.J.
      Glenoid screw position in the Encore Reverse Shoulder Prosthesis: an anatomic dissection study of screw relationship to surrounding structures.
      The compression screws are available in sizes from 18 to 46 mm in 4-mm increments. Both the standard and ExactechGPS drill bits have color-coded, etched markings on them that correspond to these increments. Regardless of navigation status, the hole depth was gauged by noting the drill bit depth marking that was flush with the drill guide at the time of cortical breakthrough. Cortical breakthrough was determined using tactile feel regardless of navigation status. However, the navigation technology also provided real-time, intraoperative animation on the navigation screen that allowed the surgeon to visualize the drill bit nearing the far cortex and the cortical breakthrough event.
      Figure thumbnail gr2
      Figure 2The ExactechGPS navigation screen is shown, demonstrating the real-time, intraoperative drill bit trajectory in relation to the patient’s scapula, rendered from a preoperative computed tomography scan.
      Superior and inferior screw positions on the baseplate were used first in all cases. The decision to place additional screws was made intraoperatively, based on the surgeon’s tactile assessment of purchase of the first 2 screws. Essentially, the quality of “bite” of the first 2 screws determined whether additional screws were used. Locking caps were then placed over each screw head, and the glenoid component was completed by placing the glenosphere according to the technique guide.
      Exactech
      Extremities operative technique: Equinoxe platform shoulder system.
      The humeral polyethylene component size was chosen based on soft-tissue tensioning; the humeral component was then secured following the technique guide. The wound was closed in standard layered fashion.

      Study design

      This study was a case-control, retrospective review of 114 rTSAs performed in 108 patients at a single institution by 1 of 2 fellowship-trained, board-certified, experienced primary surgeons. The non-navigated study group consisted of 63 surgical procedures performed from January 2018 to July 2019. Computer-assisted intraoperative navigation technology became available at our institution in October 2018. The navigated study group consisted of 51 surgical procedures performed from October 2018 to July 2019. Since navigation technology became available at our institution, 6 patients have been prevented from undergoing a navigated rTSA. This was because of either a lack of preoperative CT scans or anatomic factors that were not conducive to coracoid tracker placement, such as a small coracoid size or previous coracoid fracture. These patients were included in the non-navigated group. Preoperative surgical indications included rotator cuff arthropathy, fracture malunion, and complications from prior humeral hardware, as well as 1 chronic shoulder dislocation (listed as “other”) (Table I). The exclusion criteria included any prior major glenoid reconstruction or instrumentation, revision of prior rTSA, or conversion to rTSA.
      Table IDescriptive statistics of study population
      VariableFrequency (%)
      rTSA
       Non-navigated63 (55.3)
       Navigated51 (44.7)
      Surgery indication
       Rotator cuff arthropathy106 (93)
       Fracture3 (2.6)
       Hardware complication4 (3.5)
       Other1 (0.9)
      Walch classification
       A147 (47.5)
       A212 (12.1)
       B18 (8.1)
       B29 (9.1)
       B37 (7.1)
       C1 (1.0)
       D15 (15.2)
      Baseplate type used
       Standard63 (55.3)
       Extended cage1 (0.9)
       Posterior augmentation16 (14.0)
       Superior augmentation17 (14.9)
       Posterior-superior augmentation16 (14.0)
       Small1 (0.9)
      Total No. of screws used
       267 (59)
       325 (21.9)
       418 (15.8)
       54 (3.5)
      rTSA, reverse total shoulder arthroplasty.
      A review of the electronic medical record was performed to obtain patient age, sex, height, weight, and body mass index (BMI). A review of the operative notes was performed to record laterality of surgery, baseplate type used, number of screws used, individual screw lengths, and length of procedure. Each patient’s preoperative CT scan, if available, was then used to describe the patient’s glenoid anatomy. Glenoid retroversion was first measured from CT images using the technique described by Friedman et al
      • Friedman R.J.
      • Hawthorne K.B.
      • Genez B.M.
      The use of computerized tomography in the measurement of glenoid version.
      • Rouleau D.M.
      • Kidder J.F.
      • Pons-Villanueva J.
      • Dynamidis S.
      • Defranco M.
      • Walch G.
      Glenoid version: how to measure it? Validity of different methods in two-dimensional computed tomography scans.
      and Rouleau et al.
      • Friedman R.J.
      • Hawthorne K.B.
      • Genez B.M.
      The use of computerized tomography in the measurement of glenoid version.
      • Rouleau D.M.
      • Kidder J.F.
      • Pons-Villanueva J.
      • Dynamidis S.
      • Defranco M.
      • Walch G.
      Glenoid version: how to measure it? Validity of different methods in two-dimensional computed tomography scans.
      The posterior subluxation index was calculated from CT scans as well, using the method described by Walch et al.
      • Walch G.
      • Badet R.
      • Boulahia A.
      • Khoury A.
      Morphologic study of the glenoid in primary glenohumeral osteoarthritis.
      Each glenoid was then classified using the modified Walch classification.
      • Bercik M.J.
      • Kruse II, K.
      • Yalizis M.
      • Gauci M.O.
      • Chaoui J.
      • Walch G.
      A modification to the Walch classification of the glenoid in primary glenohumeral osteoarthritis using three-dimensional imaging.
      There were 14 patients in the non-navigated group for whom glenoid morphology could not be evaluated as they did not undergo CT scans prior to surgery. For patients with CT scans in both groups, the complexity of glenoid morphology was assessed by comparing the frequency of glenoids classified as Walch type A and those classified as non–Walch type A (Walch type B, C, or D). The composite screw length was calculated by summing all screw lengths. This measure was used as a means to quantify the total screw fixation length independent of the number of screws used.

      Statistical analysis

      SAS software (version 9.4; SAS Institute, Cary, NC, USA) was used for statistical analysis. Sample characteristics were described using descriptive statistics. Frequencies and percentages were used to describe categorical variables. Means and standard deviations or medians and ranges were used to describe continuous variables. Bivariate analysis was performed for patient-specific variables and operative variables. We performed χ2 tests for sex, laterality, Walch classification, surgeon, and baseplate type. Two-sample t tests were performed for age, posterior subluxation index, and operative time. We performed Wilcoxon rank sum tests for height, weight, BMI, glenoid retroversion, number of screws used, individual screw length, composite screw length, and frequency of 2 screws used in total. The significance level was set to P < .05 for all tests.

      Results

      Frequencies and percentages of categorical variables are provided in Table I. Bivariate comparisons of patient-specific variables are presented in Table II. The navigated and non-navigated groups showed statistically significant differences in median glenoid retroversion (4.1° [interquartile range, 47.1°] vs. 8.4° [interquartile range, 46.7°], P = .016) and the mean posterior subluxation index (0.50 ± 0.1 vs. 0.57 ± 0.1, P = .002). The 2 groups did not have statistically different glenoid morphology by the Walch classification (P = .743). Moreover, there were no other statistically significant differences between the groups for patient-specific variables, including age, height, weight, BMI, sex, and laterality.
      Table IIBivariate comparison of patient-specific variables in non-navigated and navigated rTSA groups
      Non-navigated rTSA (n = 63)Navigated rTSA (n = 51)P value
      Mean ± SD or frequency (%)Median (interquartile range)Mean ± SD or frequency (%)Median (interquartile range)
      Age, yr72.1 ± 7.936 (70.6).6733
      Height, cm165.1 (48.2)15 (29.4)167.6 (33).3559
      Weight, kg87.5 (100)71.5 ± 7.981.6 (75.8).9183
      BMI, kg/m230.8 (31.6)29.2 (22.9).7442
      Sex.2634
       Male24 (52.2)22 (47.8)
       Female39 (62.9)23 (37.1)
      Laterality.1576
       Right shoulder38 (61.3)24 (38.7)
       Left shoulder25 (48.1)27 (51.9)
      Glenoid morphology
       Glenoid retroversion, °8.4 (46.7)4.1 (47.1).0166
      Statistically significant (P < .05).
       Posterior subluxation index0.57 ± 0.10.50 ± 0.1.0020
      Statistically significant (P < .05).
       Walch classification.7437
      A30 (50.9)29 (49.2)
      B, C, or D19 (47.5)21 (52.5)
      rTSA, reverse total shoulder arthroplasty; SD, standard deviation; BMI, body mass index.
      Statistically significant (P < .05).
      Bivariate comparisons of the operative variables are presented in Table III. There was no significant difference in the frequency of procedures performed by surgeon 1 vs. surgeon 2 between the groups (P = .380). The average individual screw length was significantly higher in the navigated group (36.7 mm vs. 30 mm, P < .0001). The median composite screw length was also higher in the navigated group (84 mm vs. 76 mm, P = .048), despite the use of significantly fewer screws per case (2.5 ± 0.7 vs. 2.8 ± 1, P = .047). The frequency of using 2 screws in total was significantly higher in the navigated group (35 of 51 patients [68.6%] vs. 32 of 63 controls [50.8%], P = .047). The frequency of augmented baseplate use (39 [76.5%] vs. 12 [19.1%], P < .0001) and the operative time (98.6 ± 19.5 minutes vs. 85.8 ± 18.7 minutes, P = .001) were significantly higher in the navigated group.
      Table IIIBivariate comparison of operative variables in non-navigated and navigated rTSA groups
      Non-navigated rTSANavigated rTSAP value
      Mean ± SD or frequency (%)Median (range)Mean ± SD or frequency (%)Median (range)
      Surgeon.3808
       Surgeon 149 (77.8)36 (70.6)
       Surgeon 214 (22.2)15 (29.4)
      Baseplate screws
       No. of screws used2.8 ± 12.5 ± 0.7.0472
      Statistically significant (P < .05).
       Individual screw length, mm30 (20)36.7 (22)<.0001
      Statistically significant (P < .05).
       Composite screw length, mm76 (90)84 (86).0481
      Statistically significant (P < .05).
       Frequency of 2 screws used in total32 (50.8)35 (68.6).0472
      Statistically significant (P < .05).
      Baseplate type<.0001
      Statistically significant (P < .05).
       Standard51 (81)12 (23.5)
       Augmented12 (19.1)39 (76.5)
      Operative time, min85.8 ± 18.798.6 ± 19.5.0006
      Statistically significant (P < .05).
      rTSA, reverse total shoulder arthroplasty; SD, standard deviation.
      Statistically significant (P < .05).

      Discussion

      The patient-specific variables of the 2 study groups in this report were well matched with respect to age, sex, laterality, and BMI. A statistically significant difference of 4.3° of median glenoid retroversion (8.4° vs. 4.1°, P = .017) was found in the non-navigated group. However, both groups had median retroversion within the normal range for anatomic glenoid retroversion (1°-9°), and the 2 study groups did not have statistically different glenoid morphology by the modified Walch classification (P = .744).
      • Matsumura N.
      • Ogawa K.
      • Kobayashi S.
      • Oki S.
      • Watanabe A.
      • Ikegami H.
      • et al.
      Morphologic features of humeral head and glenoid version in the normal glenohumeral joint.
      The modified Walch classification was used to characterize glenoid morphology because it is a well-known CT-based system that was most translatable when assessing the glenoid using the 3D preoperative templating software.
      A longer individual screw length was attained using intraoperative navigation technology (36.7 mm vs. 30 mm). A longer screw length (36 vs. 18 mm), specifically in the anterior and inferior screw positions, has been associated with biomechanically stronger initial fixation, but the number of screws (2 vs. 4) and angle of screw divergence (0° vs. 27°) have not.
      • Lung T.S.
      • Cruickshank D.
      • Grant H.J.
      • Rainbow M.J.
      • Bryant T.J.
      • Bicknell R.T.
      Factors contributing to glenoid baseplate micromotion in reverse shoulder arthroplasty: a biomechanical study.
      The finding of an improved screw length with navigation technology is in concordance with the results of Nashikkar et al.
      • Nashikkar P.S.
      • Scholes C.J.
      • Haber M.D.
      Role of intraoperative navigation in the fixation of the glenoid component in reverse total shoulder arthroplasty: a clinical case-control study.
      However, the improvement was isolated to the anterior and posterior screws with a much shorter maximum screw length (20 mm) in comparison to our study. This difference is likely due to surgeon-dependent variation in preoperative goals for screw position and length. Obtaining the maximum screw length may be seen as having secondary importance to placing screws into the areas of highest bone density in the scapula, particularly the base of the coracoid, the scapular pillar, and the scapular spine.
      • Humphrey C.S.
      • Kelly II, J.D.
      • Norris T.R.
      Optimizing glenosphere position and fixation in reverse shoulder arthroplasty, part two: the three-column concept.
      Nashikkar et al routinely placed screws in at least 4 of the 5 screw holes with maximal screw divergence, possibly leading to a shorter maximum attainable length for each screw.
      In contrast, the surgeons involved in this study routinely look to minimize the number of screws used. Two screws were placed in the inferior and superior positions, at a minimum. On the basis of tactile assessment of screw purchase, or bite, the surgeon decided whether to add a third or fourth screw. Fewer screws were used on average with the use of navigation technology, with 2 screws in total used more frequently. The increased frequency of the use of 2 screws in total is not biomechanical evidence of improved fixation but does reflect increased surgeon confidence in the level of fixation intraoperatively. Improved screw length and visual confirmation of intraosseous placement of the screws using the navigation technology likely contributed to this increased confidence. A longer individual screw length has been shown to provide significantly better baseplate fixation after cyclic loading, regardless of the number of screws used.
      • Roche C.
      • DiGeorgio C.
      • Yegres J.
      • VanDeven J.
      • Stroud N.
      • Flurin P.H.
      • et al.
      Impact of screw length and screw quantity on reverse total shoulder arthroplasty glenoid fixation for 2 different sizes of glenoid baseplates.
      The post hoc planning in this study alleviates potential concerns that the surgeons used fewer screws for the sole purpose of obtaining significant results. We cannot draw any biomechanical conclusions based on these data, but these results do reflect a significant change in practice at our institution because of intraoperative navigation.
      The relationship between the number of screws used and biomechanical strength in baseplate fixation is unclear, with some evidence supporting the use of more screws and some refuting their use.
      • Chebli C.
      • Huber P.
      • Watling J.
      • Bertelsen A.
      • Bicknell R.T.
      • Matsen III, F.
      Factors affecting fixation of the glenoid component of a reverse total shoulder prothesis.
      ,
      • Hoenig M.P.
      • Loeffler B.
      • Brown S.
      • Peindl R.
      • Fleischli J.
      • Connor P.
      • et al.
      Reverse glenoid component fixation: is a posterior screw necessary?.
      ,
      • Irlenbusch U.
      • Kohut G.
      Evaluation of a new baseplate in reverse total shoulder arthroplasty—comparison of biomechanical testing of stability with roentgenological follow up criteria.
      • James J.
      • Huffman K.R.
      • Werner F.W.
      • Sutton L.G.
      • Nanavati V.N.
      Does glenoid baseplate geometry affect its fixation in reverse shoulder arthroplasty?.
      • Kennon J.C.
      • Lu C.
      • McGee-Lawrence M.E.
      • Crosby L.A.
      Scapula fracture incidence in reverse total shoulder arthroplasty using screws above or below metaglene central cage: clinical and biomechanical outcomes.
      The complex interplay between the implant design and the importance of screw trajectory to biomechanical strength makes it difficult to fully evaluate the impact of using additional screws. Theoretically, using fewer screws might preserve scapular bone stock in case revision surgery is necessary in the future. Notably, the use of fewer screws in this study did not result in loss of the average composite screw length, with 8 mm of additional average composite screw length using navigation.
      Unexpectedly, we found a drastic increase in augmented baseplate use from 19.1% to 76.5% (P < .0001) in the navigated group. Preoperative templating software has been shown to increase the awareness of glenoid bone loss and the need for augmented glenoid baseplates to achieve post-implantation version of 0°-10° of retroversion for anatomic total shoulder arthroplasty.
      • Iannotti J.P.
      • Weiner S.
      • Rodriguez E.
      • Subhas N.
      • Patterson T.E.
      • Jun B.J.
      • et al.
      Three-dimensional imaging and templating improve glenoid implant positioning.
      Even so, our use of augmentations in 76.5% of navigated shoulders is much greater than that in previous studies of rTSA and anatomic total shoulder arthroplasty.
      • Barrett I.
      • Ramakrishnan A.
      • Cheung E.
      Safety and efficacy of intraoperative computer-navigated versus non-navigated shoulder arthroplasty at a tertiary referral.
      ,
      • Iannotti J.P.
      • Weiner S.
      • Rodriguez E.
      • Subhas N.
      • Patterson T.E.
      • Jun B.J.
      • et al.
      Three-dimensional imaging and templating improve glenoid implant positioning.
      The decision to use augmentation is left to surgeon discretion. True optimal post-implantation glenoid version in rTSA has yet to be defined, and long-term outcome studies are necessary to better define the indications for augmented baseplate use. A recent study by Parsons et al
      • Parsons M.
      • Greene A.
      • Polakovic S.
      • Byram I.
      • Cheung E.
      • Jones R.
      • et al.
      Assessment of surgeon variability in preoperative planning of reverse total shoulder arthroplasty: a quantitative comparison of 49 cases planned by 9 surgeons.
      detailed significant variation in the criteria for augmented baseplate utilization among 9 surgeons but found a uniform rate of overall utilization of augmentations, at 79%. The surgeons in our study attempt to implant the baseplate perpendicular to the Friedman line to achieve what is essentially neutral version. With this goal in mind, we attribute the drastic increase in the use of augmentation to the use of the 3D animated preoperative planning software. The software provides 3D feedback on the projected baseplate version relative to the Friedman line, possibly granting a better appreciation of glenoid morphology and final baseplate version. The use of this technology and augmented baseplates has been shown to allow for correction to neutral version in 63% of cases.
      • Parsons M.
      • Greene A.
      • Polakovic S.
      • Byram I.
      • Cheung E.
      • Jones R.
      • et al.
      Assessment of surgeon variability in preoperative planning of reverse total shoulder arthroplasty: a quantitative comparison of 49 cases planned by 9 surgeons.
      The surgeons in our study anecdotally found that the use of augmentation allowed them to rely less on reaming to achieve their desired version. However, the amount of required reaming and the accuracy with which the implanted baseplate’s version matched the preoperative plan were not assessed. Multiple studies have confirmed that simply using 3D technology platforms to virtually implant glenoids preoperatively is helpful in improving surgeon accuracy in baseplate positioning intraoperatively.
      • Barrett I.
      • Ramakrishnan A.
      • Cheung E.
      Safety and efficacy of intraoperative computer-navigated versus non-navigated shoulder arthroplasty at a tertiary referral.
      ,
      • Berhouet J.
      • Gulotta L.V.
      • Dines D.M.
      • Craig E.
      • Warren R.F.
      • Choi D.
      • et al.
      Preoperative planning for accurate glenoid component positioning in reverse shoulder arthroplasty.
      ,
      • Iannotti J.P.
      • Weiner S.
      • Rodriguez E.
      • Subhas N.
      • Patterson T.E.
      • Jun B.J.
      • et al.
      Three-dimensional imaging and templating improve glenoid implant positioning.
      ,
      • Wang A.W.
      • Hayes A.
      • Gibbons R.
      • Mackie K.E.
      Computer navigation of the glenoid component in reverse total shoulder arthroplasty: a clinical trial to evaluate the learning curve.
      ,
      • Werner B.S.
      • Hudek R.
      • Burkhart K.J.
      • Gohlke F.
      The influence of three-dimensional planning on decision-making in total shoulder arthroplasty.
      The use of navigation increased the operative time by 12.8 minutes (98.6 ± 19.5 minutes vs. 85.8 ± 18.7 minutes, P = .001). The extra steps of coracoid dissection, coracoid tracker placement, and glenoid registration with the stylus likely contributed to this increase in operative time. However, the time was likely inflated by the learning curve associated with the use of new instrumentation and a new technique. Prior studies using the same system and a similar technique documented 1.2- and 6-minute increases in operative time.
      • Barrett I.
      • Ramakrishnan A.
      • Cheung E.
      Safety and efficacy of intraoperative computer-navigated versus non-navigated shoulder arthroplasty at a tertiary referral.
      ,
      • Wang A.W.
      • Hayes A.
      • Gibbons R.
      • Mackie K.E.
      Computer navigation of the glenoid component in reverse total shoulder arthroplasty: a clinical trial to evaluate the learning curve.
      The decreased intraoperative time cited in these studies likely better represents our current practice, post-learning curve.
      There are a number of limitations to our study. Because of the lack of preoperative CT scans, 14 patients in the non-navigated group were not able to be included in the analysis of glenoid morphology. There were differences between the 2 cohorts in terms of glenoid retroversion and the posterior subluxation index. Although the differences were statistically significant, they were unlikely to be clinically significant. The finding of an increased individual screw length with use of navigation must be interpreted with caution, as we were unable to assess the actual intraosseous length of the screws without subjecting patients to additional CT scans. We did not quantify screw length per screw position in our study, which may be of biomechanical relevance. This study does not provide biomechanical evidence of increased glenoid baseplate stability or accuracy of final implant position using augmented baseplates. However, our results do reflect the impact that navigation technology has had on the practice of the surgeons involved in this study. Long-term follow-up is the next logical step in our investigation. Further studies will examine clinical outcomes and early loosening rates associated with navigated rTSA.

      Conclusion

      Intraoperative computer navigation allowed for a longer individual screw length, an increased composite screw length, fewer total screws used, and an increased frequency of 2 screws used in total. Preoperative templating software led to a drastic increase in augmented glenoid baseplate use. The role of navigation and preoperative templating in rTSA is still evolving, but their use in this study significantly impacted surgical planning and the intraoperative technique.

      Disclaimer

      The work performed in this study was completed without funding from outside sources. Publication fees were provided by Exactech Incorporated. The authors, their immediate families, and any research foundations with which they are affiliated have not received any financial payments or other benefits from any commercial entity related to the subject of this article.

      References

        • Barrett I.
        • Ramakrishnan A.
        • Cheung E.
        Safety and efficacy of intraoperative computer-navigated versus non-navigated shoulder arthroplasty at a tertiary referral.
        Orthop Clin North Am. 2019; 50: 95-101https://doi.org/10.1016/j.ocl.2018.08.004
        • Bercik M.J.
        • Kruse II, K.
        • Yalizis M.
        • Gauci M.O.
        • Chaoui J.
        • Walch G.
        A modification to the Walch classification of the glenoid in primary glenohumeral osteoarthritis using three-dimensional imaging.
        J Shoulder Elbow Surg. 2016; 25: 1601-1606https://doi.org/10.1016/j.jse.2016.03.010
        • Berhouet J.
        • Gulotta L.V.
        • Dines D.M.
        • Craig E.
        • Warren R.F.
        • Choi D.
        • et al.
        Preoperative planning for accurate glenoid component positioning in reverse shoulder arthroplasty.
        Orthop Traumatol Surg Res. 2017; 103: 407-413https://doi.org/10.1016/j.otsr.2016.12.019
        • Bitzer A.
        • Rojas J.
        • Patten I.S.
        • Joseph J.
        • McFarland E.G.
        Incidence and risk factors for aseptic baseplate loosening of reverse total shoulder arthroplasty.
        J Shoulder Elbow Surg. 2018; 27: 2145-2152https://doi.org/10.1016/j.jse.2018.05.034
        • Chebli C.
        • Huber P.
        • Watling J.
        • Bertelsen A.
        • Bicknell R.T.
        • Matsen III, F.
        Factors affecting fixation of the glenoid component of a reverse total shoulder prothesis.
        J Shoulder Elbow Surg. 2008; 17: 323-327https://doi.org/10.1016/j.jse.2007.07.015
        • Exactech
        Extremities operative technique: Equinoxe platform shoulder system.
        2017 (Accessed August 19, 2019)
        • Frankle M.
        • Siegal S.
        • Pupello D.
        • Saleem A.
        • Mighell M.
        • Vasey M.
        The reverse shoulder prosthesis for glenohumeral arthritis associated with severe rotator cuff deficiency. A minimum two-year follow-up study of sixty patients.
        J Bone Joint Surg Am. 2005; 87: 1697-1705https://doi.org/10.2106/JBJS.D.02813
        • Friedman R.J.
        • Hawthorne K.B.
        • Genez B.M.
        The use of computerized tomography in the measurement of glenoid version.
        J Bone Joint Surg Am. 1992; 74: 1032-1037
        • Hart N.D.
        • Clark J.C.
        • Wade Krause F.R.
        • Kissenberth M.J.
        • Bragg W.E.
        • Hawkins R.J.
        Glenoid screw position in the Encore Reverse Shoulder Prosthesis: an anatomic dissection study of screw relationship to surrounding structures.
        J Shoulder Elbow Surg. 2013; 22: 814-820https://doi.org/10.1016/j.jse.2012.08.013
        • Hoenig M.P.
        • Loeffler B.
        • Brown S.
        • Peindl R.
        • Fleischli J.
        • Connor P.
        • et al.
        Reverse glenoid component fixation: is a posterior screw necessary?.
        J Shoulder Elbow Surg. 2010; 19: 544-549https://doi.org/10.1016/j.jse.2009.10.006
        • Humphrey C.S.
        • Kelly II, J.D.
        • Norris T.R.
        Optimizing glenosphere position and fixation in reverse shoulder arthroplasty, part two: the three-column concept.
        J Shoulder Elbow Surg. 2008; 17: 595-601https://doi.org/10.1016/j.jse.2008.05.038
        • Iannotti J.P.
        • Weiner S.
        • Rodriguez E.
        • Subhas N.
        • Patterson T.E.
        • Jun B.J.
        • et al.
        Three-dimensional imaging and templating improve glenoid implant positioning.
        J Bone Joint Surg Am. 2015; 97: 651-658https://doi.org/10.2106/JBJS.N.00493
        • Irlenbusch U.
        • Kohut G.
        Evaluation of a new baseplate in reverse total shoulder arthroplasty—comparison of biomechanical testing of stability with roentgenological follow up criteria.
        Orthop Traumatol Surg Res. 2015; 101: 185-190https://doi.org/10.1016/j.otsr.2014.11.015
        • James J.
        • Huffman K.R.
        • Werner F.W.
        • Sutton L.G.
        • Nanavati V.N.
        Does glenoid baseplate geometry affect its fixation in reverse shoulder arthroplasty?.
        J Shoulder Elbow Surg. 2012; 21: 917-924https://doi.org/10.1016/j.jse.2011.04.017
        • Kennon J.C.
        • Lu C.
        • McGee-Lawrence M.E.
        • Crosby L.A.
        Scapula fracture incidence in reverse total shoulder arthroplasty using screws above or below metaglene central cage: clinical and biomechanical outcomes.
        J Shoulder Elbow Surg. 2017; 26: 1023-1030https://doi.org/10.1016/j.jse.2016.10.018
        • Königshausen M.
        • Jettkant B.
        • Sverdlova N.
        • Ehlert C.
        • Gessmann J.
        • Schildhauer T.A.
        • et al.
        Influence of different peg length in glenoid bone loss: a biomechanical analysis regarding primary stability of the glenoid baseplate in reverse shoulder arthroplasty.
        Technol Health Care. 2015; 23: 855-869https://doi.org/10.3233/THC-151031
        • Lung T.S.
        • Cruickshank D.
        • Grant H.J.
        • Rainbow M.J.
        • Bryant T.J.
        • Bicknell R.T.
        Factors contributing to glenoid baseplate micromotion in reverse shoulder arthroplasty: a biomechanical study.
        J Shoulder Elbow Surg. 2019; 28: 648-653https://doi.org/10.1016/j.jse.2018.09.012
        • Matsumura N.
        • Ogawa K.
        • Kobayashi S.
        • Oki S.
        • Watanabe A.
        • Ikegami H.
        • et al.
        Morphologic features of humeral head and glenoid version in the normal glenohumeral joint.
        J Shoulder Elbow Surg. 2014; 23: 1724-1730https://doi.org/10.1016/j.jse.2014.02.020
        • Nashikkar P.S.
        • Scholes C.J.
        • Haber M.D.
        Role of intraoperative navigation in the fixation of the glenoid component in reverse total shoulder arthroplasty: a clinical case-control study.
        J Shoulder Elbow Surg. 2019; 28: 1685-1691https://doi.org/10.1016/j.jse.2019.03.013
        • Nguyen D.
        • Ferreira L.M.
        • Brownhill J.R.
        • King G.J.
        • Drosdowech D.S.
        • Faber K.J.
        • et al.
        Improved accuracy of computer assisted glenoid implantation in total shoulder arthroplasty: an in-vitro randomized controlled trial.
        J Shoulder Elbow Surg. 2009; 18: 907-914https://doi.org/10.1016/j.jse.2009.02.022
        • Parsons M.
        • Greene A.
        • Polakovic S.
        • Byram I.
        • Cheung E.
        • Jones R.
        • et al.
        Assessment of surgeon variability in preoperative planning of reverse total shoulder arthroplasty: a quantitative comparison of 49 cases planned by 9 surgeons.
        J Shoulder Elbow Surg. 2020; 29: 2080-2088https://doi.org/10.1016/j.jse.2020.02.023
        • Roche C.
        • DiGeorgio C.
        • Yegres J.
        • VanDeven J.
        • Stroud N.
        • Flurin P.H.
        • et al.
        Impact of screw length and screw quantity on reverse total shoulder arthroplasty glenoid fixation for 2 different sizes of glenoid baseplates.
        JSES Open Access. 2019; 3: 296-303https://doi.org/10.1016/j.jses.2019.08.006
        • Rouleau D.M.
        • Kidder J.F.
        • Pons-Villanueva J.
        • Dynamidis S.
        • Defranco M.
        • Walch G.
        Glenoid version: how to measure it? Validity of different methods in two-dimensional computed tomography scans.
        J Shoulder Elbow Surg. 2010; 19: 1230-1237https://doi.org/10.1016/j.jse.2010.01.027
        • Routman H.D.
        • Flurin P.H.
        • Wright T.W.
        • Zuckerman J.D.
        • Hamilton M.A.
        • Roche C.P.
        Reverse shoulder arthroplasty prosthesis design classification system.
        Bull Hosp Jt Dis (2013). 2015; 73: S5-S14
        • Venne G.
        • Rasquinha B.J.
        • Pichora D.
        • Ellis R.E.
        • Bicknell R.
        Comparing conventional and computer-assisted surgery baseplate and screw placement in reverse shoulder arthroplasty.
        J Shoulder Elbow Surg. 2015; 24: 1112-1119https://doi.org/10.1016/j.jse.2014.10.012
        • Verborgt O.
        • De Smedt T.
        • Vanhees M.
        • Clockaerts S.
        • Parizel P.M.
        • Van Glabbeek F.
        Accuracy of placement of the glenoid component in reversed shoulder arthroplasty with and without navigation.
        J Shoulder Elbow Surg. 2011; 20: 21-26https://doi.org/10.1016/j.jse.2010.07.014
        • Walch G.
        • Badet R.
        • Boulahia A.
        • Khoury A.
        Morphologic study of the glenoid in primary glenohumeral osteoarthritis.
        J Arthroplasty. 1999; 14: 756-760
        • Wang A.W.
        • Hayes A.
        • Gibbons R.
        • Mackie K.E.
        Computer navigation of the glenoid component in reverse total shoulder arthroplasty: a clinical trial to evaluate the learning curve.
        J Shoulder Elbow Surg. 2020; 29: 617-623https://doi.org/10.1016/j.jse.2019.08.012
        • Werner B.S.
        • Hudek R.
        • Burkhart K.J.
        • Gohlke F.
        The influence of three-dimensional planning on decision-making in total shoulder arthroplasty.
        J Shoulder Elbow Surg. 2017; 26: 1477-1483https://doi.org/10.1016/j.jse.2017.01.006
        • Zumstein M.A.
        • Pinedo M.
        • Old J.
        • Boileau P.
        Problems, complications, reoperations, and revisions in reverse total shoulder arthroplasty: a systematic review.
        J Shoulder Elbow Surg. 2011; 20: 146-157https://doi.org/10.1016/j.jse.2010.08.001